Catalog Number 397003-001 |
Device Problems
Electrical /Electronic Property Problem (1198); Image Display Error/Artifact (1304)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because it did not prevent the companion 2 driver from performing its life-sustaining functions.The companion hospital cart will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer, a syncardia certified hospital, reported that the monitor on the hospital cart stopped working while supporting a patient.
|
|
Manufacturer Narrative
|
The customer-reported experience of the hospital cart display that "stopped working" could not be confirmed or reproduced during the investigation testing.Visual inspection of the hospital cart did not reveal any damage or abnormalities, and the cart passed all functional testing, which includes display functionality.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4730 follow-up report 1.
|
|