Catalog Number 0620040610 |
Device Problem
Protective Measures Problem (3015)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/20/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that there was an error message for a high pressure reading during surgery, indicating overpressure.
|
|
Event Description
|
It was reported that there was an error message for a high pressure reading during surgery, indicating overpressure.
|
|
Manufacturer Narrative
|
Alleged failure: cib kk asr please eval error messages on system during case.Po sjc the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.There is no probable root cause in relation to the complaint as there were no issues observed during the testing of the insufflator.However, the unit is of an older revision and has components that are outdated.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|