• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Catalog Number 0165L14
Device Problems Deflation Problem (1149); Device Dislodged or Dislocated (2923); Device Fell (4014)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 03/10/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the catheter fell out spontaneously.
 
Manufacturer Narrative
The reported event was unconfirmed since the problem could not be reproduced.The evaluation did not find any leakage observed even when pressure was applied on sac.The sample went under a 7 day leak test and no leakage was observed.There was no leakage from the valve.The catheter was dissected and no conditions were found that could have contributed to the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿(1)do not reuse.(2)do not resterilize.(3)this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4)be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5)do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients (1)do not use in patients who are or have been allergic to natural rubber latex.".
 
Event Description
It was reported that the catheter fell out spontaneously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8525393
MDR Text Key142613369
Report Number1018233-2019-02034
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2022
Device Catalogue Number0165L14
Device Lot Number8NP030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2019
Date Manufacturer Received05/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-