BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Model Number 9770602Y |
Device Problems
No Display/Image (1183); Defective Component (2292)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device will not be returned to the manufacturer for evaluation, as the device is being retained by the reporting facility.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per nurse manager: the unit has a raining effect in the image.Originally the rain was only on the low contrast and #2 gain.Now it has become more random and is happening more frequently on multiple settings and there have been 3 times when the probe did not work at all.Meaning there was no image, no rain, just fuzzy white lines across the machine.That was resolved by restarting the ultrasound several times.We (facility) exchanged the probe on the original machine with one that we (facility) have not had any difficulty with and it worked well with that probe.Then we (facility) put the bad probe on the good machine and had the same "rain" effect on the good machine."this is why we think it is probably a problem with the probe.".
|
|