• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770602Y
Device Problems No Display/Image (1183); Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned to the manufacturer for evaluation, as the device is being retained by the reporting facility.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per nurse manager: the unit has a raining effect in the image.Originally the rain was only on the low contrast and #2 gain.Now it has become more random and is happening more frequently on multiple settings and there have been 3 times when the probe did not work at all.Meaning there was no image, no rain, just fuzzy white lines across the machine.That was resolved by restarting the ultrasound several times.We (facility) exchanged the probe on the original machine with one that we (facility) have not had any difficulty with and it worked well with that probe.Then we (facility) put the bad probe on the good machine and had the same "rain" effect on the good machine."this is why we think it is probably a problem with the probe.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND SYSTEM 20MM PINPOINT GT PROBE (REMANUFACTURED)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP.
141 zehner school road
zelienople PA 16063
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key8528427
MDR Text Key142634988
Report Number3006260740-2019-00962
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138201
UDI-Public(01)00801741138201
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770602Y
Device Catalogue Number9770602Y
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 03/27/2019
Initial Date FDA Received04/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-