• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; SCREW, LOCKING BONE, RSP, 5X14

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; SCREW, LOCKING BONE, RSP, 5X14 Back to Search Results
Model Number 506-03-114
Device Problem Difficult to Remove (1528)
Patient Problem No Code Available (3191)
Event Date 03/21/2019
Event Type  Injury  
Event Description
Revision surgery - due to a screw needing to be removed.Reason was not given by surgeon.
 
Manufacturer Narrative
The reason for this revision surgery was to remove a screw, reason unknown.The previous surgery and the surgery detailed in this investigation occurred 3 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery to remove a screw, reason unknown.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
SCREW, LOCKING BONE, RSP, 5X14
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key8528603
MDR Text Key142395892
Report Number1644408-2019-00356
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024532
UDI-Public(01)00888912024532
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number506-03-114
Device Catalogue Number506-03-114
Device Lot Number831C1583
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/21/2019
Initial Date FDA Received04/18/2019
Supplement Dates Manufacturer Received04/18/2019
Supplement Dates FDA Received05/06/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
508-02-032, LOT 983W1015; 508-32-103, LOT 864C3262
Patient Outcome(s) Required Intervention;
-
-