CAREFUSION, 303 INC. BD ALARIS PUMP FUSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2426-0007
Device Problems Device Displays Incorrect Message (2591); Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/15/2019
Event Type  malfunction  
Event Description
Pump tubing was primed, and placed into pump to infuse to the pt.The clamps were open when infusion started.After medication infusing into pt's picc, the pump started to alarm for "channel error".The rn opened the pump channel door to find a balloon effect had developed in top tubing segment that gets locked into the channel.Pump and channel used for pt care were evaluated by clinical engineering.No defects or concerns found from the pump malfunctioning.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS PUMP FUSION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION, 303 INC.
MDR Report Key8531541
MDR Text Key142734613
Report NumberMW5085995
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/19/2022
Device Model Number2426-0007
Device Catalogue Number2424-0007
Device Lot Number(10) 19016336
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/18/2019
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Weight81
-
-