HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC Back to Search Results
Model Number P3700E000057
Device Problem Device Slipped (1584)
Patient Problem No Patient Involvement (2645)
Event Date 04/10/2019
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the brake detent needed to be replaced.Per the hill-rom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in february 2019.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the brake detent to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the hill-rom service technician stating the brakes would not hold.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
TABLE, OBSTETRIC
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key8535805
MDR Text Key145537699
Report Number1824206-2019-00171
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberP3700E000057
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/10/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-