The assignable cause for the higher than expected vitros ft4 result and lower than expected vitros tsh result could not be determined with the information provided.The customer did not provide any qc fluid information, so it was not possible to make an appropriate assessment of the vitros reagent performance at the time of the event for either assay.Additionally, precision testing of the vitros system was not performed, therefore it cannot be confirmed that the instrument was operating as intended and unexpected instrument performance cannot be completely ruled out as contributing to the events.In addition, it was not possible to establish if either customer was following the sample collection device manufacturer¿s recommended centrifugation protocol; therefore, pre-analytical sample processing cannot be ruled out as a contributing factor.It is possible that cellular debris, due to poor sample preparation, was present in the affected samples, although this could not be confirmed.Furthermore, the sample was reported to be visibly icteric and the sample may have interfered with the vitros assays.No sample remained for investigative testing at the time the issue was reported.
|