• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. DRILL PIN AND SCREW INSERTER; EXTRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. DRILL PIN AND SCREW INSERTER; EXTRACTOR Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a knee arthroplasty procedure, the instrument tip fractured during screw insertion all the piece were returned.No adverse event or delay was reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of drill pin and screw inserter shows signs of repeated use (nicked or gouged) and it is fractured.Device history record was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to routine wear and tear after nearly eight years in the field.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL PIN AND SCREW INSERTER
Type of Device
EXTRACTOR
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8544028
MDR Text Key142890454
Report Number0001822565-2019-01697
Device Sequence Number1
Product Code HWB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 11/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00590102100
Device Lot Number61787275
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/27/2019
Initial Date FDA Received04/24/2019
Supplement Dates Manufacturer Received11/08/2019
Supplement Dates FDA Received11/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-