• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558380
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/28/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, there was a strong bend in the preloaded guidewire within the balloon when inflating and deflating.The procedure was completed with this device.There were no patient complications reported as a result of this event.
 
Event Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, there was a strong bend in the preloaded guidewire within the balloon when inflating and deflating.The procedure was completed with this device.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block h6: problem code 2981 captures the reportable issue of distal tip bent.Block h10: investigation results: a visual examination of the returned complaint device found that the catheter was kinked in two different sections.No other issues were noted on the returned device.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Type of Device
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8544109
MDR Text Key142901561
Report Number3005099803-2019-02055
Device Sequence Number1
Product Code KNQ
UDI-Device Identifier08714729202011
UDI-Public08714729202011
Combination Product (y/n)N
PMA/PMN Number
K971320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/21/2021
Device Model NumberM00558380
Device Catalogue Number5838
Device Lot Number0023226372
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2019
Initial Date Manufacturer Received 03/29/2019
Initial Date FDA Received04/24/2019
Supplement Dates Manufacturer Received05/22/2019
Supplement Dates FDA Received06/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-