• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUEST MEDICAL, INC. MPS DELIVERY SET; CPBP HEAT EXCHANGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUEST MEDICAL, INC. MPS DELIVERY SET; CPBP HEAT EXCHANGER Back to Search Results
Model Number 5001102
Device Problems Failure to Deliver (2338); Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/04/2019
Event Type  malfunction  
Manufacturer Narrative
The device will be evaluated and a follow up medwatch will be submitted if additional information becomes available.
 
Event Description
A report was received regarding an alleged issue encountered during use of the mps delivery set.The report states that during a case, the mps displayed inadequate fill.Arterial flow was high but clamp would not open.The console was changed out and they used the same disposable without further issues but when they tried to give cardioplegia, the mps would not deliver.They took the cross clamp off, changed the disposable, gave cardioplegia and the case proceeded without further issues.There were no reported patient complications resulting from the alleged incident.
 
Manufacturer Narrative
The device was received and the complaint condition was confirmed.The tubing on the blood side of the pump cassette was assembled too far into the bushing and this resulted in the "inadequate fill" alarm as reported.A quality alert was issued as a corrective action.Quest will continue to monitor trends for the reported complaint condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPS DELIVERY SET
Type of Device
CPBP HEAT EXCHANGER
Manufacturer (Section D)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
Manufacturer (Section G)
QUEST MEDICAL, INC.
one allentown parkway
allen TX 75002
Manufacturer Contact
tosan onosode
one allentown parkway
allen, TX 75002
9723326338
MDR Report Key8545215
MDR Text Key146360769
Report Number1649914-2019-00024
Device Sequence Number1
Product Code DTR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173716
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5001102
Device Lot Number0576899F10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2019
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-