• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Death (1802); Cardiac Perforation (2513)
Event Date 04/11/2019
Event Type  Death  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the right inferior pulmonary vein (ripv) was perforated.Additionally, the physician noticed blood in the endotracheal (et) tube.Blood was suctioned from the et tube, but bleeding continued, and the case had to be aborted while the patient was under general anesthesia.The patient is deceased.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed eight applications were performed with balloon catheter 2af284 with lot number 96020 without any issues present.A clinical issue (cardiac perforation and death) was encountered during the procedure.The case was aborted while the patient was under general anesthesia.In conclusion, the reported sheath was not returned for investigation.There is no indication of relation of the adverse event to the performance of the cryo device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8547205
MDR Text Key143018218
Report Number3002648230-2019-00296
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00763000043773
UDI-Public00763000043773
Combination Product (y/n)N
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/15/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number88766
Was Device Available for Evaluation? No
Date Manufacturer Received04/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
2AF284 BALLOON CATHETER, 990063-015 MAPPING CATH
Patient Outcome(s) Death; Required Intervention;
Patient Age76 YR
-
-