Model Number SN60WF
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The cartridge product history record could not be reviewed because facility did not provide a lot number or any identification traceable to the manufacturing documentation.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
 
Event Description
A facility representative is reporting an issue with an intraocular lens (iol), with a description of "lens was deformed".There was no patient contact and the procedure was completed the same day.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key8555155
MDR Text Key143535529
Report Number1119421-2019-00551
Device Sequence Number1
Product Code HQL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.205
Device Lot Number12639385
Was Device Available for Evaluation? No
Date Manufacturer Received04/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-