It was reported that during the implant procedure, while testing the left ventricular (lv) threshold, there was no capture at high output.I look at the fluoroscopy noted the lead had pulled back slightly.The physician disconnected the lv lead and tried put a guidewire through the lead to provide stability.The wire would not advance in the inner lune of the lead very far.Physician noted blood in or on the insulation of the lead where the wire would not pass.Physician then flushed the lead with saline using a sub-dermal needle and noted that the saline was leaking out where the wire was getting held up.The physician cut the lead just proximal to the apparent insulation issue.The physician reinserted the guidewire through the lead was able to successfully place it into the coronary sinus (cs) and maintain venous access while removing the lead lv lead.It was also noted there possible fracture.The lead lv lead was attempted and not used.A different lv lead was used to complete the procedure.No patient complications have been reported as a result of this event.
|
Product event summary: the full lead was returned in segments, analyzed.Analysis indicated that the distal conductor was extrinsically distorted due to kinking/buckling.The proximal conductor of the lead became extrinsically distorted due to kinking/buckling.The outer insulation of the lead was extrinsically breached due to a cut.Visual analysis of the lead indicated damage during use.The analyst noted that visual inspection found the conductors were kinked, and the insulation was cuts.It looks likely that conductor kink occured when the anchoring sleeve tied down during implant.After the lead was placed and dislodged, the insulation could be cut during re-position the lead.According to the reported and the analysis results, it is likely that after the lead was placed and dislodged, when try to re-position the lead, the guidewire could not go through because the conductors were kinked, and the solution saline was leaking out due to the insulation breached.If information is provided in the future, a supplemental report will be issued.
|