• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date of event: only event year known: 2019.Batch # r92812.Device analysis: the analysis results of the el5ml found that the device was received with the jaw broken and the jaw was not returned for analysis.The device was disassembled and evidence of corrosion was found throughout the broken area.3 remaining clips were found on the clip track.The most likely reason for jaw breakage is stress corrosion cracking and the most likely root cause is exposure to a solution containing chlorine.The reported complaint was confirmed.In addition, in order to avoid this kind of issues please do not reuse, reprocess or re sterilize device.Reuse, reprocessing or re sterilization may compromise the structural integrity of the device and/or lead to device failure that in turn may result in patient injury, illness or death.A manufacturing record evaluation was performed for the finished device batch r92812 number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the clip was not advancing.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8559992
MDR Text Key143869854
Report Number3005075853-2019-18528
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 03/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2019
Initial Date Manufacturer Received 04/01/2019
Initial Date FDA Received04/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-