• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/14/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t9288d.Device analysis: the analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled, upon disassembly the advancer was confirmed to be bent and 6 clips were found inside clip track.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device lot/batch number t9288d, and no non-conformances were identified.
 
Event Description
It was reported that during a sleeve gastrectomy the device was closed on tissue and fired but the clips did not advance and deploy.The procedure was completed with an alternate like device with no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8560266
MDR Text Key143877235
Report Number3005075853-2019-18534
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Catalogue NumberEL5ML
Device Lot NumberT9288D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2019
Date Manufacturer Received04/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-