Model Number 2202-0007 |
Device Problems
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
Patient Problem
No Code Available (3191)
|
Event Date 03/21/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical product: 150ml baxter bag container lot dr18k12045.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
It was reported that upon entering the infant's room to hang platelets, the nurse discovered a small white puddle on the floor that was determined to be 20% intralipids.The rate of infusion was 4ml/hr to infuse over 20 hours using a 1.2 micron filter.Upon further investigation, the nurse noticed that the lipids where slowly dripping down the pump module.When the nurse opened the channel door a tear in the iv tubing was found to be leaking the lipids.There was "emotional distress or inconvenience" for the patient.
|
|
Manufacturer Narrative
|
The customer¿s report that the tubing leaked and had a tear in the iv tubing was not confirmed.Inspection under magnification of the set's components and infusion bag observed no obvious damages or issues.An attempted reprime, infusion ran at the reported rate, and pressure tested of the received set noted no leaks or any issues.Further testing was performed.A lab gauge needle was manually attached to the male luer end of the reprimed set and the set was loaded into a lab pump module.An infusion was programmed to run at the reported rate of 4ml for one hour.The infusion completed as expected with no leaks, alarms, or any issues observed.The set was also pressure tested while submerged underwater.Each of the set's tubing engagements were manipulated as well as the silicone segment component being slightly stretched while the pressure was gradually increased up to 30psi.No leaks or any issues were observed.The root cause was not identified.
|
|
Event Description
|
It was reported that upon entering the infant's room to hang platelets, the nurse discovered a small white puddle on the floor that was determined to be 20% intralipids.The rate of infusion was 4ml/hr to infuse over 20 hours using a 1.2 micron filter.Upon further investigation, the nurse noticed that the lipids where slowly dripping down the pump module.When the nurse opened the channel door a tear in the iv tubing was found to be leaking the lipids.
|
|
Search Alerts/Recalls
|