Model Number 397002-001 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver will be evaluated at syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver system check results were outside the testing acceptance criteria.
|
|
Manufacturer Narrative
|
The root cause of the reported issue, system check results outside test acceptance criteria, was determined to be malfunctions of the right and left pilot valves.Syncardia has a corrective and preventive action (capa) for issue related to pilot valve performance.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4739 follow-up report 1.
|
|