Model Number 397002-001
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 04/09/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver will be evaluated at syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver system check results were outside the testing acceptance criteria.
 
Manufacturer Narrative
The root cause of the reported issue, system check results outside test acceptance criteria, was determined to be malfunctions of the right and left pilot valves.Syncardia has a corrective and preventive action (capa) for issue related to pilot valve performance.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4739 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8564708
MDR Text Key146384685
Report Number3003761017-2019-00082
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/09/2019
Initial Date FDA Received04/30/2019
Supplement Dates Manufacturer Received04/09/2019
Supplement Dates FDA Received07/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-