The cause for the discordant total hcg results is unknown.Siemens healthcare diagnostics is investigating.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with the patient's medical history, clinical presentation, and other findings." the ifu states in the limitations section: "all in vitro assays can generate erroneous results, both clinically false positive results (test results suggesting a condition that is absent) and clinically false negative results (test results failing to identify a condition that is present).".
|
Siemens filed the initial mdr on may 3, 2019.06/04/2019 additional information: one patient has shown discordant results with two different samples, drawn on different days: (b)(6) was tested on (b)(6) 2019, result was negative (<0.05 miu/ml).(b)(6) was tested on (b)(6) 2019, result was negative (<0.05 miu/ml).(b)(6) was tested on (b)(6) 2019, result was 42 miu/ml.(b)(6) was tested on (b)(6) 2019, result was 3.5 miu/ml.(b)(6) was tested on (b)(6) 2019, result was negative (<0.05 miu/ml).Siemens reviewed the data, there was no service performed on the instrument.In this case the discordant result was due to gaining rlu for a direct assay.There are multiple potential causes for this issue (wash/ aspiration issues or acid/base delivery).The review of the instrument files did not reveal a chronic instrument issue or a product problem.Based on this investigation no product problem was identified.The cause for the discordant total hcg result is unknown.The instrument is operational.No further evaluation of the device is required.
|