• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 MP T1 PPS SO 5X95MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 MP T1 PPS SO 5X95MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Problem with Sterilization (1596)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/23/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that there was debris in the sterile packaging.No adverse events have been reported as a result of the malfunction.Additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  complaint sample was evaluated and the reported event was confirmed.The returned product was visually inspected and the reported event (debris in packaging) was confirmed.Device history record was reviewed and no discrepancies were found.The likely condition of the product when it left zimmer biomet control was non-conforming.The root cause of the reported event is the operator not following instructions during the manufacturing process.This product falls within the scope of a corrective action which is reviewing issues with debris in packaging.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 MP T1 PPS SO 5X95MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8582098
MDR Text Key144133000
Report Number0001825034-2019-02045
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K110400
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 08/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-108050
Device Lot Number6429590
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/23/2019
Initial Date FDA Received05/06/2019
Supplement Dates Manufacturer Received08/12/2019
Supplement Dates FDA Received08/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-