MAUDE Adverse Event Report: MICROTYPING SYSTEMS BLOOD GROUPING REAGENT MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD; MTS GEL CARDS - BLOOD GROUPING REAGENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROTYPING SYSTEMS BLOOD GROUPING REAGENT MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD; MTS GEL CARDS - BLOOD GROUPING REAGENTS Back to Search Results
Catalog Number MTS080515
Device Problem False Positive Result (1227)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/22/2019
Event Type  malfunction  
Manufacturer Narrative
Ortho performed retain testing, batch review, complaint review by lot and master lot.Based upon the results of this investigation, the reported customer issue was unable to be confirmed.All results were satisfactory.Sample was not returned to ortho for further investigation.(b)(4).
 
Event Description
Report 1 of 2.One patient with a previous history of rh negative using tube typing, typed rh positive on the vision using mts card lot# 072318037-68.(site contact not aware if weak d testing was performed) account had another sample drawn and testing demonstrated the same false positive results.Account sent the sample to the arc who also performed testing in tube with weak d testing and determined the sample to be rh negative.No harm came to the patient.Account will only selected rh negative cells for possible transfusion.Reported (b)(6) 2019.Microtubes/wells or cell (donor #) affected: d microtube.Reaction grade obtained: 2+.Customer was expecting: negative.Incubation time (for manual test only): na.Test repeated: yes - on a new sample and by sending sample to a reference lab.Method/result obtained by repeating: new sample at this lab - still 2+.Reference lab negative including weak d testing.Daily qc performed and found to be acceptable.Product handling protocol: cassette/gel card storage temperature range: per ifu.Cassette/gel card orientation: upright.Rbc storage and handling: per ifu.Visual appearance before use: all reagents have a normal appearance prior to use.Opened vs unopened? opened.Tsc: informed account about the differences in product clones and how this may account for the discrepancy.In addition, inadequate centrifugation of the sample may cause the presence of fibrin and/or cellular debris.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD GROUPING REAGENT MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD
Type of Device
MTS GEL CARDS - BLOOD GROUPING REAGENTS
Manufacturer (Section D)
MICROTYPING SYSTEMS
1295 southwest 29th avenue
pompano beach FL 33069
Manufacturer (Section G)
MICROTYPING SYSTEMS
1295 southwest 29th avenue
pompano beach FL 33069
Manufacturer Contact
matthew p wictome
1001 route 202
raritan, NJ 08869
9082188223
MDR Report Key8582348
MDR Text Key201891694
Report Number1056600-2019-00017
Device Sequence Number1
Product Code KSZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date07/03/2019
Device Catalogue NumberMTS080515
Device Lot Number072318037-68
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2019
Initial Date FDA Received05/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-