No device was returned for examination.Photography and video evidence suggests the wire was unraveling during the cutting application.In the video provided, it is noticed that the wire is unraveling as they are cutting.That wire in the video has many crimp marks so the wire used was in poor condition contributing to the unraveling that was occurring.It has been determined that to perform the cut test correctly, the cable must be new and the cutting motion has to be swift.The intended clinical application is to have the cable under tension.However, as the device was not returned a functional test could not be performed.The complaint is unable to be confirmed.Lot identification is necessary for review of device history records, lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|