Model Number SEL-00002
Device Problem Unintended System Motion (1430)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/16/2019
Event Type  malfunction  
Event Description
It was reported that the rotation switch on the detector was not working correctly, sometimes it does not stop moving because the switch is blocked.A field engineer was dispatched to the site and it was determined that bracket from the switch was bent.Once the bracket was adjusted the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key8583169
MDR Text Key144168884
Report Number1220984-2019-00043
Device Sequence Number1
Product Code MUE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSEL-00002
Device Catalogue NumberSEL-00002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-