• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BHR ACETABLR CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BHR ACETABLR CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Device Problems Corroded (1131); Insufficient Information (3190)
Patient Problems Toxicity (2333); Injury (2348); No Code Available (3191)
Event Date 09/13/2017
Event Type  Injury  
Event Description
It was reported that right hip revision surgery was performed due to metallosis, pseudotumor and highly elevated cobalt and chromium levels.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the bhr cup and bhr head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the head and cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the cup.Similar complaints were identified for head and this will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.Although the reported pain, elevated ions and intraoperative findings of necrosis and pseudotumor may be consistent with findings associated with metallosis the root cause cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR ACETABLR CUP
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8583271
MDR Text Key144167326
Report Number3005975929-2019-00194
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 04/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Date Manufacturer Received03/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FEMORAL HEAD, PART AND LOT # UNKNOWN; FEMORAL HEAD, PART AND LOT # UNKNOWN
Patient Outcome(s) Hospitalization; Required Intervention;
-
-