• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty transformer on the processor/bridge/pace board.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that during a routine shift check by a clinician, the device failed self-test for pacer function.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8584276
MDR Text Key144245622
Report Number1220908-2019-01284
Device Sequence Number1
Product Code DRO
UDI-Device Identifier00847946005979
UDI-Public00847946005979
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-