It was reported that friction was encountered during the advancement of the 3rd coil in a subarachnoid hemorrhage anterior communicating artery aneurysm, and the coil got stuck inside the catheter.During removal, the coil detached in the catheter.There was no reported patient injury or intervention.The patient was reported to be doing well.
|
The pusher, microcatheter, and implant were received for evaluation; a visual analysis revealed the implant was not attached to the pusher.The proximal hypotube of the pusher was bent.The distal body coil of the pusher was damaged.The heater coil of the pusher did not contain any indication of thermal damage.The microcatheter underwent x-ray examination.The implant was noted to be within the lumen of the catheter, with the proximal end of the implant at the strain relief portion of the catheter.The implant was stretched throughout the catheter, until the distal end of the coil was seen.The microcatheter does not have any indication of being smashed.The distal black pet covering the heater coil was removed.The heater coil looked to be intact, and the solder segments were normal.The attachment monofilament was dissected out of the pusher, and the tip of the monofilament was noted to be stretched.Based on the evaluation of the returned device, the reported complaint was confirmed, as the implant was returned detached from the pusher and stuck inside the microcatheter.The root cause for the implant becoming stuck could not be determined, as x-ray does not show any smashed segment damage on the catheter, or any clumping of the implant that created an oversized od.The root cause of the detachment could not be definitively determined; however, the implant most likely detached due to experiencing over specification of force.
|