• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY; URINE DRAINAGE BAG (CONOD)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY; URINE DRAINAGE BAG (CONOD) Back to Search Results
Catalog Number 319114A
Device Problems Device Contamination with Chemical or Other Material (2944); Device Contaminated During Manufacture or Shipping (2969)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/11/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that a white foreign material was found inside the inlet tube 2 hours after placement.
 
Event Description
It was reported that a white foreign material was found inside the inlet tube 2 hours after placement.
 
Manufacturer Narrative
The reported event was confirmed.The evaluation observed a yellowish stain inside the inlet tube.How and when event are occurred could not be determine.A potential root cause for this failure mode could be defect contributed by vendor or customer/ improper handling/ unclean machine /working area.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[contraindications]: 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients (1) patients who are or have been allergic to natural rubber latex (2) patients with known allergy to silver coated catheter".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY
Type of Device
URINE DRAINAGE BAG (CONOD)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8585929
MDR Text Key144260998
Report Number1018233-2019-02337
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2022
Device Catalogue Number319114A
Device Lot NumberMYCY1062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2019
Date Manufacturer Received06/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-