Model Number LNQ11 |
Device Problems
Incorrect Measurement (1383); Unable to Obtain Readings (1516)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/19/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the implantable cardiac monitor (icm) experienced invalid data and no ecg view during symptoms episodes due to a "flipped bit" in the icm memory.The icm remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The implantable cardiac monitor (icm) is no longer in use.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.The returned device indicated an issue in an integrated circuit.Hybrid analysis confirmed the reported patient assistant egm field failure.The fault was isolated to the stacked chip scale package.If information is provided in the future, a supplemental report will be issued.
|
|