• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Date 04/05/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate and could confirm the reported issue.He replaced the push-button switch.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The affected part was returned to livanova (b)(4) for a detailed investigation.The investigator could reproduce the reported issue and found that the switch did not meet its datasheet.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the switch of a s5 roller pump did not work properly during setup.There was no patient involvement.
 
Manufacturer Narrative
The replaced switch was returned to livanova deutschland for further investigation.During the evaluation the reported issue could be reproduced.The switch was found out of specification.Livanova deutschland notified the supplier about the reported issue and requested further actions to prevent recurrence.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial.
 
Event Description
See initial.
 
Manufacturer Narrative
The address in the section has been corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8586674
MDR Text Key145975501
Report Number9611109-2019-00327
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number10-80-00
Was the Report Sent to FDA? No
Date Manufacturer Received05/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-