Medcomp received the adverse incident report directly from the (b)(6) in the (b)(6) and immediately contacted our distributor in the (b)(6) and asked them to contact the facility to obtain additional information and the sample if available.At this time the device is unknown.The lot number provided on the report, mlb8580, is not accurate for a medcomp product.
|
A femoral catheter was inserted for hemodialysis access.During the insertion procedure, the catheter and dilators were flushed for patency.With the aid of uss guidance, the vein was punctured in a single attempt with good flows on aspiration.The procedure went through smoothly with cannulation of the vein, and guidewire was inserted under uss guidance.Some resistance was encountered which required repositioning of the needle for smooth insertion of the guidewire.Position of the guidewire into the vein was confirmed on ultrasound.Dilator was passed over the guidewire with full visibility of the wire.The catheter was placed over the guidewire with the wire protruding/visible from the blue lumen.The catheter was inserted successfully and the guidewire was extracted with mild resistance.Confirming good aspiration from both lumens the catheter was sutured in place and patient was moved for dialysis.Reviewing the patient and catheter on dialysis, it was noted that a metallic object was protruding from the blue end of the catheter.The sutures were cut and the catheter position was manipulated, which helped to push the guidewire further towards the surface where it was then pulled out.A section of the guidewire was extracted from the femoral catheter.The end of the wire was curved which confirmed this was the end point of the guidewire.The patient remained well during and after the procedure.
|