It was reported that during the treatment of an anterior communicating aneurysm, the coil was advanced out approximately 2cm.During removal, resistance was encountered, and the coil detached in the aneurysm.The coil was successfully removed with a snare device.Several attempts were made to clarify if the coil was thermally detached or detached; however, no further clarifying information has been provided.There was no reported patient injury.The patient was reported to be fine.
|
The device was received with the dispenser hoop, introducer sheath, delivery pusher, and the implant, which was detached from the delivery pusher.The heater coil of the pusher did not have any burn damage, but did display compression damage.The rest of the pusher did not display any other notable damage.The implant had multiple kinks.The attachment monofilament displayed a tail shape.The root cause of the implant detaching is most likely due to the monofilament experiencing a force over the maximum specification, as indicated by the physical characteristics of the monofilament tip.
|