• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD ANALYZER; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN EMERALD ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 09H59-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
 
Event Description
The customer observed a falsely elevated wbc result on the cell-dyn emerald analyzer.The following data was provided: 1st sample: cd emerald (4am sample) wbc 97.7 k/l, beckman lh: 408 k/l, alinity h: 407 k/l.2nd sample recollected sample at 7am same day: cd emerald: 100 k/l, beckman lh: 395 k/l, alinity h: 407 k/l.There was no impact to patient management reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of historical data and a review of labeling.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN EMERALD ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key8593815
MDR Text Key169925934
Report Number2919069-2019-00061
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740020286
UDI-Public00380740020286
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09H59-01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/22/2019
Initial Date FDA Received05/08/2019
Supplement Dates Manufacturer Received06/17/2019
Supplement Dates FDA Received06/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-