The device with the lens was returned.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger has been advanced over the lens to mid-nozzle.The lens is still partially in the loading area under the plunger.The trailing haptic is folded in on the optic.The nozzle was removed and cleaned for further evaluation.The lens was removed during cleaning.Top coat dye stain testing was conducted with acceptable results.A qualified viscoelastic was indicated.The root cause for the plunger override could not be determined.The plunger was advanced over the lens, the plunger was not advanced through the lens.The plunger position advanced to mid-nozzle with the lens still in the loading area may indicate the plunger was advanced too rapidly.Additional information was provided in d.10., h.3., h.6.And h.10.The manufacturer internal reference number is: (b)(4).
|