Unique identifier: (b)(4).Suspect medical device: there are no additional device identification numbers.Device evaluated by mfr: the investigation by ge healthcare (gehc) has been completed.A gehc field engineer completed system testing and verified that the mri scanner was operating normally and within performance specifications.The root cause was determined to be operator error related to the use of a 3rd party medtronic axiem non-invasive patient tracker (nipt) during the patient mri scan.Per discussion with the user, the medtronic mr conditions for use were not strictly followed during this procedure, resulting in an rf burn to the patient.Per our gehc operator manual, the user must ensure the safe use of mr conditional devices during patient scans.Users should consult the device manufacturer's instructions and safety guidelines.No further actions are planned by gehc.Medtronic has been notified of the patient event involving their product.
|