MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG RT SM SIZE 4 PMA; KNEE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG RT SM SIZE 4 PMA; KNEE PROSTHESIS Back to Search Results
Model Number N/A
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Pain (1994)
Event Date 04/09/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: medical product: oxf uni tib tray sza rm/ll pma, catalog #: 154719, lot # 056950: medical product: oxf twin-peg cmntd fem sm pma, catalog #: 161468, lot #: 441720.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2019-00446 3002806535-2019-00447.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial right knee procedure.Subsequently, the patient was revised due to pain and loosening.
 
Manufacturer Narrative
(b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial right knee procedure.Subsequently, the patient was revised due to pain and loosening.Information received that the femoral and tibial components were well fixed, the bearing was popping out.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that a patient underwent an initial right knee procedure.Subsequently, the patient was revised due to pain and loosening.Information received stated that the femoral and tibial components were well fixed, the bearing was popping out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG RT SM SIZE 4 PMA
Type of Device
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
MDR Report Key8598547
MDR Text Key144661799
Report Number3002806535-2019-00448
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup,Followup
Report Date 09/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2017
Device Model NumberN/A
Device Catalogue Number159569
Device Lot Number2762535
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight170
-
-