ZOLL CIRCULATION AUTOPULSE® PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Display Difficult to Read (1181)
Patient Problem No Patient Involvement (2645)
Event Date 04/17/2019
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of a blank screen on the autopulse platform during power on was confirmed during functional testing.The investigation findings revealed that the lcd was missing pixels.Therefore, the root cause of the reported blank screen was due to a defective lcd as result of wear and tear.The returned autopulse platform (serial # (b)(4)) was manufactured in february 2011 and it is 8 years old, well beyond the expected serviceable life of 5 years.Other than the missing pixels on the lcd, there was no physical damage on the autopulse platform was observed during visual inspection.Initial functional testing could not performed due to platform's blank screen.To remedy the reported issue, the defective lcd identified during service evaluation was replaced.During re-evaluation, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test battery until discharged without any fault or error.The ap platform passed all functional tests and it is ready for clinical use.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaints reported for autopulse with serial number (b)(4).
 
Event Description
During shift check, customer reported that the autopulse platform (serial # (b)(4)) does power on but the display screen was blank.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key8602105
MDR Text Key144977137
Report Number3010617000-2019-00374
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000772
UDI-Public00849111000772
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-