With respect to the returned unit it has passed all specific functional testing requirements, except for the lock having rotational movement when unit is not under pressure, this would not have caused a slippage; when unit is properly positioned and put under pressure unit would not have slipped.Unit needed heli-coils added to large starburst threads.The device was manufactured in 2012.The reported complaint was not confirmed.Root cause was unknown however the most probable root causes are: incorrectly assembled device, improperly tested/inspected device, improper technique used during procedure, off-label use of device during procedure (user error), device used with incorrect/unauthorized devices/accessories for procedure.
|