MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Device Problems Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/15/2019
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the balloon catheter, 2af284 with lot number 56618, was returned and analyzed.Visual inspection of the balloon catheter showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used for 4 injections.The catheter passed the performance test and electrical integrity as per specification; but dissection and pressure test revealed a guide wire lumen kink at 1.02 inches from the tip inside the balloons.In conclusion, the balloon catheter failed the returned product inspection due to a guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the balloon catheter subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8604520
MDR Text Key144851532
Report Number3002648230-2019-00353
Device Sequence Number1
Product Code OAE
UDI-Device Identifier00643169939318
UDI-Public00643169939318
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/06/2019
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number56618
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2019
Initial Date Manufacturer Received 04/17/2019
Initial Date FDA Received05/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-