During an atrial fibrillation ablation procedure the patient experienced hypotension and a transient ischemic attack (tia).While ablating, the patient's atrial fibrillation converted, the underlying rhythm was sinus arrest with junctional escape and the patient developed associated hypotension.Intracardiac pacing was done to stabilize the patient during the procedure; however, the patient was hypotensive post procedure and required chronotropic and inotropic support in the icu.The patient's blood pressure was continually monitored following the procedure until a stable blood pressure was reached.Additionally, 50 minutes following ablation the patient was presenting with left-sided facial droop.A ct brain scan confirmed right middle cerebral artery (mca) stroke, with filling defect of the terminal right mca branches in the distal m3 region.During the procedure the active clotting time (act) was approximately 350 seconds and was maintained with heparin to reach the target act, transesophageal echo was performed prior to the start of the procedure to confirm no left atrial appendage thrombus was present.Clot retrieval was not an option and the following morning all neurological abnormalities resolved.The hypotension and tia resolved without sequela.
|
The results of the investigation are inconclusive since the device was not returned for analysis; however, the log files from the tactisys quartz system were returned.The log file analysis concluded that the tacticath catheter performed as intended.The optical fibers met specifications, the recorded temperatures indicated cooling during rf ablation, and contact force measurements were displayed throughout the duration of the log files.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported hypotension and transient ischemic attack remains unknown.
|