• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Pacing Problem (1439)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a faulty capacitor on the pace/defib engine board.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Event Description
Complainant alleged that during biomed testing, the device's pacer output was out of specification.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8609491
MDR Text Key145027104
Report Number1220908-2019-01277
Device Sequence Number1
Product Code DRO
UDI-Device Identifier00847946017200
UDI-Public00847946017200
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 04/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/29/2019
Initial Date FDA Received05/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-