• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Electrical /Electronic Property Problem (1198); Audible Prompt/Feedback Problem (4020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/22/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the companion 2 driver continued to perform its life-sustaining functions.The companion 2 driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the companion 2 driver did not recognize external batteries anymore and exhibited a battery error alarm while supporting a patient.The customer also reported that the patient was switched to a backup driver and there was no reported adverse patient impact.
 
Manufacturer Narrative
The patient data file and alarm history was reviewed and revealed battery-related alarms that could have been caused either from lack of recognition (depleted batteries) or from the batteries being removed from the driver.Investigation testing determined that the driver passed all requirements and functioned as intended.Additionally, external batteries that have previously met performance requirements were used to test the ability of the driver to recognize and charge external batteries, which the driver was able to do.The customer-reported issue of the driver not detecting external batteries was not able to be reproduced during investigation testing.The driver performed as intended with no evidence of a device malfunction.The external batteries used at the time of the customer-reported issue were not returned and therefore could not be evaluated as part of this investigation.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4764 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8610253
MDR Text Key145112604
Report Number3003761017-2019-00102
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2019
Initial Date Manufacturer Received 04/24/2019
Initial Date FDA Received05/14/2019
Supplement Dates Manufacturer Received04/24/2019
Supplement Dates FDA Received06/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age48 YR
-
-