Model Number 397002-001 |
Device Problems
Electrical /Electronic Property Problem (1198); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the companion 2 driver continued to perform its life-sustaining functions.The companion 2 driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia authorized distributor, reported that the companion 2 driver did not recognize external batteries anymore and exhibited a battery error alarm while supporting a patient.The customer also reported that the patient was switched to a backup driver and there was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The patient data file and alarm history was reviewed and revealed battery-related alarms that could have been caused either from lack of recognition (depleted batteries) or from the batteries being removed from the driver.Investigation testing determined that the driver passed all requirements and functioned as intended.Additionally, external batteries that have previously met performance requirements were used to test the ability of the driver to recognize and charge external batteries, which the driver was able to do.The customer-reported issue of the driver not detecting external batteries was not able to be reproduced during investigation testing.The driver performed as intended with no evidence of a device malfunction.The external batteries used at the time of the customer-reported issue were not returned and therefore could not be evaluated as part of this investigation.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4764 follow-up report 1.
|
|
Search Alerts/Recalls
|