• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/22/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the issue did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver is also equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and an external battery charger which provide additional means of mitigating the impact to patient care in the event that the power adapter cannot be connected.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The connection 1 receptacle is the connector located on the freedom driver housing that connects the power adaptor to the freedom driver.The customer, a syncardia authorized distributor, reported that the freedom power adaptor plug was broken at the freedom driver connection 1 receptacle while supporting a patient.The customer also reported that the patient was switched to the backup freedom driver with no adverse patient impact.
 
Manufacturer Narrative
The customer-reported broken connection to the power adaptor was confirmed via visual inspection where it was observed that the connection 1 receptacle cable was displaced in the driver.This is the connection point between the power adaptor and the driver.It was no longer attached to the driver housing and was lodged inside the housing.This confirmed the customer-reported issue.It is unknown how the connection 1 receptacle was damaged, but is likely the result of either the application of excessive force when connecting the power adaptor to the driver, or rough handling / impact shock to the driver.Despite the damaged connection 1 receptacle cable, the driver passed all functional testing requirements.Syncardia has a corrective and preventive action (capa) to investigate broken connection 1 receptacle cables.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4762 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8610257
MDR Text Key145540072
Report Number3003761017-2019-00100
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 07/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2019
Date Manufacturer Received04/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age43 YR
-
-