Model Number 595000-001 |
Device Problems
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because the issue did not prevent the freedom driver from performing its life-sustaining functions.The freedom driver is also equipped with redundant power sources, including multiple rechargeable freedom onboard batteries and an external battery charger which provide additional means of mitigating the impact to patient care in the event that the power adapter cannot be connected.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The connection 1 receptacle is the connector located on the freedom driver housing that connects the power adaptor to the freedom driver.The customer, a syncardia authorized distributor, reported that the freedom power adaptor plug was broken at the freedom driver connection 1 receptacle while supporting a patient.The customer also reported that the patient was switched to the backup freedom driver with no adverse patient impact.
|
|
Manufacturer Narrative
|
The customer-reported broken connection to the power adaptor was confirmed via visual inspection where it was observed that the connection 1 receptacle cable was displaced in the driver.This is the connection point between the power adaptor and the driver.It was no longer attached to the driver housing and was lodged inside the housing.This confirmed the customer-reported issue.It is unknown how the connection 1 receptacle was damaged, but is likely the result of either the application of excessive force when connecting the power adaptor to the driver, or rough handling / impact shock to the driver.Despite the damaged connection 1 receptacle cable, the driver passed all functional testing requirements.Syncardia has a corrective and preventive action (capa) to investigate broken connection 1 receptacle cables.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4762 follow-up report 1.
|
|
Search Alerts/Recalls
|