PHARMACY LITE PACKAGING 33MM GREEN EZ ONE VIALS; CONTAINER, LIQUID MEDICATION GRADUATED Back to Search Results
Model Number GEZ-150
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2019
Event Type  malfunction  
Event Description
We had a parent of an individual that we fill liquid prescription for bring up an issue they were having with the gez-150 bottles that we dispense the liquid preparation in.We typically do 360ml of this particular suspension in each gez-150 bottle.The parent's complaint was that on two separate occasions, once when the bottle was nearly full and once when it was about 1/4 full, when they are drawing a dose out of the bottle they happened to drop the vial on the laminate floor in their home and the bottle shattered (not just cracked but shattered producing pieces of plastic).They estimate a 5 ft fall height, bottles being stored refrigerated (35-45 degrees fahrenheit - nothing frozen) and dose being taken out immediately after shaking suspension while the bottle is still cold.A majority of our products that we dispense in these bottles require refrigeration, and though dropping a bottle is not an ideal situation, should they not stand up to that kind of impact? is this a part of the company's quality control process for these bottles? is there something regarding the plastic's chemical make up of the particular batches that we received? we were a little skeptical at first, but then we tried it in our pharmacy.We filled a bottle full of water (which may potentially be lighter than the suspension) and simulated the situation at both room temperature and refrigerated (35-45 degrees f).The bottle held up to the impact at room temperature, but when cold the bottle broke open near the base and produced plastic shards.We have purchased a total of 6 boxes, with 4 full boxes on hand and one currently in use.Emphasis on bottle being cold and breaking if dropped.Fda safety report id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
33MM GREEN EZ ONE VIALS
Type of Device
CONTAINER, LIQUID MEDICATION GRADUATED
Manufacturer (Section D)
PHARMACY LITE PACKAGING
MDR Report Key8611952
MDR Text Key145486015
Report NumberMW5086634
Device Sequence Number1
Product Code KYW
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 05/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberGEZ-150
Device Catalogue Number0222
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received05/14/2019
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age6 YR
-
-