(b)(4).This report is being submitted late, due to an internal retrospective review.Product has been received by zimmer biomet and the investigation is complete.Complaint sample was evaluated and the reported event was confirmed.It is possible that the reamer was damaged due to the off-axis use in the tunnel.Dhr was reviewed, and no anomalies were noted.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|