• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number ASK-05401-NM
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the kit contained broken vials of medication.
 
Event Description
It was reported that the kit contained broken vials of medication.
 
Manufacturer Narrative
(b)(4).The lot # (23f18l0491) provided by the customer is for kit mp-17019-tkl.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed with a potentially relevant finding.A nonconformance was initiated only as a general nc to temporarily allow certain finished good part numbers to be packaged with additional gauze in order to better secure medication ampules in the inner tray.Complaint verification testing could not be performed as no sample was returned for analysis.The lot # provided by the customer was for another kit.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed with a potentially relevant finding.Based on the information provided, the potential root cause of this issue is design related.A capa was initiated to further investigate this complaint issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8612794
MDR Text Key145245360
Report Number1036844-2019-00638
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASK-05401-NM
Device Lot Number23F18L0491
Was Device Available for Evaluation? No
Date Manufacturer Received06/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-