Model Number 9770500 |
Device Problems
Image Display Error/Artifact (1304); Poor Quality Image (1408)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received by the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed, the nurses claim that the image resolution is dark and grainy.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of poor image quality was unconfirmed.The scanner was evaluated with the model 550 phantom and the image was found to be normal for a sr 8 scanner.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed, the nurses claim that the image resolution is dark and grainy.
|
|