Model Number 9770501 |
Device Problem
Charging Problem (2892)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been received by the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per tm, unit will not hold a charge.'not holding a charge to the point where they cannot complete a picc if the machine is not plugged in'.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the unit will not hold a charge is confirmed.The scanner shuts down after ac power is removed.The root cause of the reported issue is an internal li-ion battery failure.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per tm, unit will not hold a charge.'not holding a charge to the point where they cannot complete a picc if the machine is not plugged in'.
|
|