DENTSPLY VDW GMBH VDW SILVER RECIPROCATING MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041163000000
Device Problem Calibration Problem (2890)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Calibration related issues in endo motors have led to file separation in the past.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply (b)(4).Micro motor (b)(4) (resistance to high)defective,replaced with (b)(4).Foot control (b)(4) and contra angle (b)(4) without errors.
 
Event Description
In this event it was reported that a vdw.Silver recirpoc device is not working properly.The investigation found that the calibration resistance is too high.The event outcome is unknown as of this mdr evaluation.
 
Manufacturer Narrative
Multiple unsuccessful attempts were made to obtain the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW SILVER RECIPROCATING MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
MDR Report Key8617539
MDR Text Key145398208
Report Number9611053-2019-00092
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2019
Date Manufacturer Received07/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-