• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE MECHANICAL LITHOTRIPTOR V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE MECHANICAL LITHOTRIPTOR V Back to Search Results
Model Number BML-V442QR-30
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/01/2019
Event Type  Injury  
Manufacturer Narrative
The subject device referenced in this report has not yet been returned to olympus for evaluation.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event Description
During an endoscopic mechanical lithotripsy, the subject device was used.In the procedure, the pipe inside the insertion portion of the device broke off.The user use the emergency lithotriptor handle to crush the stone.The basket wire of the device broke off at the proximal side.Distal side part of the wire remained in the patient's body.The patient was carried out a surgery.This is the report regarding the switching to the surgery.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that during crushing the calculus, a larger load than durability strength of the product was applied due to the factors such as size, hardness, and shape of the calculus.As a result, the basket wire and the junction between the operating pipe and the operating wire might be broken.The above device handling has warned in the instruction manual as follows.*do not use this instrument for a calculus that is assumed impossible to be crushed by a lithotripter.The pipe or the basket wire may break and part of this instrument may remain in the body.*use this instrument by having the settings to switch to open surgery and the hospitalization plan ready in case the calculus cannot be crushed by lithotripter bml-110a-1.*operation of this instrument is based on the assumption that open surgery is possible as an emergency measure.If the calculus is too hard, it is possible that the damages shown in chapter 5, ¿emergency treatment¿ may occur.Use this instrument by considering that it may lead to damage of the instrument and that open surgery may have to take place.*this instrument will deform and/or deteriorate by performing lithotripsy.When lithotripsy is repeated, it will deform and/or deteriorate furthermore.By such deformation and/or deterioration, calculus may not be crushed and/or the instrument with calculus engaged may not be removed from the body.If lithotripsy is required to be repeated in a single case, make sure to check each time that no abnormality is found in action and/or appearance (e.G.Basket wire cut or worn, tube sheath bent, notable coil sheath bent or gap etc.).Stop use when any abnormality is detected.*during lithotripsy, keep the portion from the coil sheath to the bml handle straight in line with the scope¿s biopsy valve, as much as possible.If not straight, the coil sheath may bend, calculus may not be crushed, and/or the instrument with calculus engaged may not be removed from the body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE MECHANICAL LITHOTRIPTOR V
Type of Device
SINGLE USE MECHANICAL LITHOTRIPTOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8618487
MDR Text Key145349983
Report Number8010047-2019-01964
Device Sequence Number1
Product Code LQC
UDI-Device Identifier04953170218422
UDI-Public04953170218422
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 08/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberBML-V442QR-30
Device Lot Number73K
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/01/2019
Initial Date FDA Received05/16/2019
Supplement Dates Manufacturer Received08/06/2019
Supplement Dates FDA Received08/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-