• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71940-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Itching Sensation (1943); Skin Irritation (2076); Burning Sensation (2146)
Event Date 04/29/2019
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported experiencing an adverse skin reaction after wearing the adc freestyle libre for 9 days, with symptoms of itching and burning.The customer further reported having had contact with an hcp who prescribed unspecified "cortisone-based cream" for treatment.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Since no product has been returned, extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.The reported complaint is related to skin irritation and allergic reactions to the patch adhesive of the libre sensor.Dose audit and environmental monitoring reports were reviewed for issues relating to sterility of the product.The sensor component has no impact on product sterility and therefore, sensor component dhrs (device history review) were not reviewed.Sensor kit dhrs have been reviewed to assess the manufacturing process, which includes the application of the adhesive to the puck.The dhr for the libre sensor kit was reviewed and showed the libre sensorkit passed all tests prior to release.Dose audit reports were reviewed and demonstrated the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.
 
Event Description
A customer reported experiencing an adverse skin reaction after wearing the adc freestyle libre for 9 days, with symptoms of itching and burning.The customer further reported having had contact with an hcp who prescribed unspecified "cortisone-based cream" for treatment.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key8620479
MDR Text Key145395923
Report Number2954323-2019-04012
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71940-01
Was Device Available for Evaluation? No
Date Manufacturer Received09/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient Weight95
-
-